1. Color OLED 320x240 Pixels Display
2. Live Simultaneous Color A-Scan and Thickness Display
3. Control of Gain, Blanking, Gate, Range, Delay, RF and Rectify Modes
4. Time-based B-Scan Display
5. Echo-Echo Mode (Thru Paint & Coatings)
6. Stores 100,000 Readings and 1000 waveform
If the ordinary thickness gauges are incompetent in the rough applications. Please turn to our UM-5, which will maximize measurement performance for challenging applications and provide you with the cost-effective measurement solutions.
UM-5DL Operation Principle
The ordinary ultrasonic thickness gauge with the principle pulse/echo method could do the test by the following two conditions.
1. The first underside echo must be higher than the gate. (The height of the electrical level of the gate is unadjustable)
2. No other noise waves before the first underside echo are higher than the gate. (Otherwise, we will get the thickness where the noise wave are produced)
For some conditions, the above situations could not be satisfied,for example, the highly corrosive near surface, coarse-grain materials (e.g. cast iron),aluminum, small diameter pipes, super thin sheet, super thick plate, rough surface, nonuniform inner structure, defect contained workpiece, lamination and so on which could not be tested by the ordinary gauges.
UM-5DL will solve the above problems easily:
1. Make the first underside wave higher than the gate by adjusting the gain and height of the gate.
2. Invalidate noise waves before the first underside wave by the function of blanking.
MAIN FEATURES OF UM-5DL
Live Color A-Scan
User could directly see the color waveform of the ultrasonic sound (or A-scan) on the screen, which is quite important for the occasions that we need to check the correctness of the testing results. Many cases will cause wrong testing results or even no readings. We could find the causes easily through the A-scan. Adjust the three parameters of GAIN, BLACKING, GATE, and then we will get the right readings.The uses for waveform: testing the correctness of thickness readings, finding out the causes of impacting the testing, and adjusting the parameters to solve the problems.
GAIN
Adjust the amplification factor of echo signal, increasing and reducing by 1 dB by hand. It is quite effective for acoustic attenuation materials (e.g. cast iron).
Specifications
Pulse echo with dual element transducers
Measuring Range
0.5 to 508mm(0.02" to 20.00"),
depending on probe,material and surface condition
Unit and Display Resolution
mm-0.01, 0.1
inch-0.001, 0.01
Display Mode
Normal
A-Scan
B-Scan
Max./Min.Capture
DIFF/RR%
V- Path Correction
Automatic
Measuring Error
±0.05mm (25mm below)
±0.2% H (100mm above)
±0.5% H (100mm above)
H is the height of testing object
Repeatability
±0.05mm
Display
2.4" color OLED, 320 X 240 pixels
Measurement Update Rate
Selectable4Hz, 8Hz, 16Hz
|
Velocity Range
500-9999m/s, 0.0197-0.3937in/us
Operating Language
Chinese,English,Japanese (Selectable)
Power Supply
Two 1.5V AA batteries
Operating Time
Two 1.5V AA batteries, Approximately 36 hours.
Auto Shut-off
After 5,10,20 minutes of non-action period
Working Temperature
-10ºC-+50ºC, available for as low as -20ºC on request
Size
153mm×76mm×37mm (H×W×D)
Weight
280g including batteries
Standard Package
UM-5DL thickness gauge unit
Standard Probe (TC510)
Carrying Case
Couplant (Empty Bottle)
Operating Mannual
Optional Accessories
Numerous special probes; step calibration;
Couplant and Hight temperature Couplant
|
Data Logger Option Features
|
Capacity
|
100,000 readings, 1000 waveforms,400 files
|
File Structure
|
Grid file
|
Rows X Columns
|
21 X 12
|
Communication Port
|
USB 2.0 (Full Speed ) port
|
Software
|
DataView PC software
|
UM-5 Series Probe/Transducer Specifications
|
Model
|
PT08
|
TC510
|
PT12
|
ZT12
|
PT06
|
PT04
|
GT12
|
Type
|
UM-5
Standard
|
UM-5D/DL Standard
|
Standard
|
Cast iron
|
Small tube
|
Fingertip
|
High-Temperature
|
Frequency
|
5MHz
|
5MHz
|
5MHz
|
2MHz
|
7.5MHz
|
10MHz
|
5MHz
|
Contact Diameter
|
11mm
|
13.5mm
|
13.5mm
|
17mm
|
8mm
|
6mm
|
15mm
|
Measurement Range
|
0.8 to
100.0mm
|
1.2 to
200.0mm
|
1.0 to 200.0mm
|
4.0 to
508.0mm
|
0.8 to
30.0mm
|
0.7 to 12.0mm
|
4.0 to
80.0mm
|
Temperature Range
|
-10 to 70ºC
|
-10 to
70ºC
|
-10 to 70ºC
|
-10 to 70ºC
|
-10 to
70ºC
|
-10 to 70ºC
|
-20 to 300ºC
|